Due to software malfunction, numerical values may be rounded resulting in inaccurate measurements.
Routine quality control testing of affected blood culture identification panel with affected control panel may need to be amended, due to control panel/instrument characteristic changes, that lead C. tropicalis to be amplified in some portion of test runs, but then it's reported as "Not Detected" because the assay amplicon Tm is outside the acceptable limit assigned for calling positive results.
Potential hardware issue for ARTIS icono ceiling system: collision sensor may falsely signal non-existing collisions for up to half an hour when the system is switched off for a longer time (e.g., overnight) and switched on. During this time, system movement is only possible with the "override" function with very slow speed and without further collision protection. This may lead to temporary unavailability of the system for diagnostic and therapeutic usage.