Potential packaging issue that could compromise the sterile barrier. Defect may result in loss of sterility of the device, presenting potential risk of infection if device is implanted.
Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).
Potential packaging issue that could compromise the sterile barrier. Defect may result in loss of sterility of the device, presenting potential risk of infection if device is implanted.