Loading...
Loading...
All product recalls affecting Alabama.
Includes nationwide recalls (tagged “Nationwide”) and recalls with Alabama in the distribution area.
Potential for inaccurate fetal heart rate measurements when monitoring multiples.
Apollo Tires (US) Inc. (Apollo Tires) is recalling certain Vredestein Hitrac, Quatrac, and Quatrac Pro+ tires. A contaminated compound may cause the steel belts to separate.
When a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS) and is running with the system software versions 1.17.0 and below, the analyzer does not send a message to the LAS when the sample wheel is full, and the LAS continues sending samples despite the limited capacity to process new samples. Therefore, sample tests may be aliquoted but not processed when the number of tests associated with onboard patient samples exceeds the throughput limit. If this issue occurs test results are flagged with a sample wheel timeout (SWT) error flag, and there may be a delay in reporting test results which could lead to a delay in patient treatment.
Their is a potential that the implantable pulse generator replacement indicator and end of service date may be shorter than indicated in the product labeling potentially resulting in loss of therapy.
Blind screws of the modular tibial component cannot be loosened intraoperatively,prolongation surgery due to intraoperatively change in procedure, probably to cementing technique
Due to older version of Instruction for Use (IFU) instructing to ensure that the value and the sheath are tightly connected before use. This may cause users to over-tightening the luer which may result in the swivel nut detaching from the HVA body.
ST Engineering Hackney, Inc. (STE Hackney) is recalling certain 2023-2024 Kidron Refrigerated Trailers. The wheel lug nuts may not be tightened properly.
Diamond Coach Corporation (Diamond Coach) is recalling certain 2023-2024 VIP 2200 transit buses, equipped with certain GO-ES Space Saver Foldaway Seats. The recliner mechanism may not engage when the seat back is fully reclined. As such, these vehicles fail to comply with the requirements of Federal Motor Vehicle Safety Standard number 210, "Seat Belt Assembly Anchorages."
Their is a potential that the implantable pulse generator replacement indicator and end of service date may be shorter than indicated in the product labeling potentially resulting in loss of therapy.
Volkswagen Group of America, Inc. (Audi) is recalling certain 2022-2024 Q4 E-TRON and Q4 E-TRON Sportback vehicles. The incorrect headlight control module software may have been installed. As such, these vehicles fail to comply with the requirements of Federal Motor Vehicle Safety Standard number 108, "Lamps, Reflective Devices, and Associated Equipment."
Failed Impurities/Degradation Specifications: failed related substance identified as Eptifibatide dimer.
Marketed Without An Approved NDA/ANDA: FDA laboratory analysis found the product to contain undeclared diclofenac and methocarbamol,
The reason for this recall is Beckman Coulter has determined that when a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS); is running with the software versions 1.17 and below; and a sample tube is sent from the LAS to DxI 9000 Access Immunoassay Analyzer, an out of sequence message will be sent from the Analyzer to the LAS. This issue may result in multiple error messages (i.e., a sample pick and place (PnP) error or a no result obtained (NRT) flag may be displayed) and subsequently a risk of delays in processing/reporting patient test results and patient treatment.
Turtle Top is recalling certain 2024 Van Terra, Terra Transit, Odyssey, and Odyssey XL vehicles, equipped with certain GO-ES Space Saver Foldaway Seats. The recliner mechanism may not engage when the seat back is fully reclined. As such, these vehicles fail to comply with the requirements of Federal Motor Vehicle Safety Standard number 210, "Seat Belt Assembly Anchorages."
Cummins, Inc. (Cummins) is recalling certain Cummins L9 diesel engines with part number D563027BX03. The remote fuel filter hoses may become separated from the fitting, which can result in a fuel leak and a lack of fuel supply to the engine.
Due to a software issue the PC unit may not connect to the server which could impact wireless data transmission to and from the server.
Prime-Time Specialty Vehicles (Prime-Time) is recalling certain 2023-2024 Ford Transit vehicles, equipped with certain GO-ES Space Saver Foldaway Seats. The recliner mechanism may not engage when the seat back is fully reclined. As such, these vehicles fail to comply with the requirements of Federal Motor Vehicle Safety Standard number 210, "Seat Belt Assembly Anchorages."
Kia America, Inc. (Kia) is recalling certain 2022 Niro EV vehicles. The high voltage battery safety plug may overheat and melt, which can cause a loss of drive power or fire.
A single batch of ISO-100 IsoLoc Gas-Release Endorectal Balloons is mislabeled. The affected ISO-100 balloon catheter has a nominal fill volume of 100cc/ml but the inner pouch identifies the product as an ISO-60 (fill volume of 60cc/ml).
VITROS Chemistry Products FS Diluent Pack 3, Lot 01-1645 may cause results to be negatively biased when used to dilute patient samples tested with VITROS Chemistry Products CRP Slides.