The Deltec, Graseby, and 3M, Large Volume Infusion Pump - Models 3000 and 3100, may deliver an unintended bolus if the pump door is opened and then immediately closed. This problem is due to a timing delay when the pumping mechanism resets itself.
The Smith Medical PM, Inc. BCI 3180 Pulse Oximeter tantalum capacitors C38, C72, C74, and C98 were installed backwards during assembly at the board manufacturer. This exposes the components to reverse polarity voltages that may lead to their pre-mature failure. As a result, the identified monitors may exhibit failure modes such as the monitor may shut down and not produce alarms or signals to
The Deltec, Graseby, and 3M, Large Volume Infusion Pump - Models 3000 and 3100, may deliver an unintended bolus if the pump door is opened and then immediately closed. This problem is due to a timing delay when the pumping mechanism resets itself.