An incorrect Magnesium calibrator 2 value (CAL 2) was provided for Clinical Chemistry Iron/Magnesium Calibrator lot number 54187M200. Use of the incorrect Magnesium CAL 2 value causes quality control and patient magnesium results to be reduced by as much as 17.24% above 0.8mEq/L (1.0 mg/dL or 0.4 mmol/L).
Flaws in strip voltage continuity may result in frequency of an Error 3 ("Er3") message display and unreadable strips. This issue has the potential for creating a delay in generating blood glucose results and to delay diabetes management, potentially leading to either hypo or hyperglycemia and their associated complications.
The plastic tip on the drive shaft may come off the shaft and cause the blade to stop oscillating.