When the ISOLOC (Version 4.2) profile distance unit is set to "centimeter" and the user clicked the "back" button in the digitizer window, the marker and target coordinates in the localization window erroneously display values ten times greater than the correct values.
The recall includes one-seater models 1136-3020, 1136-3021, 1136-3420, 1136-3421, 1205-3020 and two-seater models 2336-3021, 2336-3022, 2336-5019, 2575-3020, 2575-3420, 2605-3416, 2605-3515, 2606-3016, 2606-3017, 2606-3416, 2606-3417, 2636-3016, 2636-3017, 2906-3015, 2906-3016, 2906-3017, 2906-4015, 2906-4016, 2906-4017, 2909-3015, 2909-3415, 2909-4015, 2910-3015, 2910-3016, 2910-3415, 2910-4015, 2910-4016, 2910-4415, 2911-4015, 2911-4016, 2946-3015, and 2946-3016. The model number and serial number are located under the seat (on the driver's side if two-seater). The Fun-Karts were produced in several colors, to include red, black, blue, purple and green.
When ISOLOC version 6.0 and 4.2 distance unit is set to centimeter, the localization coordinates were rounded to the nearest millimeter which could affect positioning by 0.5 mm.