CK Total in Calibration Serum Level 3, CAL2351, is running with a positive bias on RX Series instruments compared to other methods experience a shift in Quality Control and patient sample recovery of up to 13%.
Cassette did not meet an 1n-process flow criterion.
Internal backup battery may fail spontaneously while the Atlan is being operated without AC mains supply, resulting in unexpected shutdown of the device while running on batteries.
Firm was informed of an event in which the structural integrity of the MR system components failed due to unintended excessive pressure buildup of helium gas during a magnet quench. If the system is unable to contain the pressured helium gas due to an unforeseen blockage in the system, risks to patients and/or operators that may lead to injury or death include chemical exposure (i.e. helium gas) that may expand to surrounding rooms, contusion caused by debris, and asphyxia.
Packaging of the device may cause a low seal strength in the side seal of the foil pouch potentially resulting in a sterile/moisture barrier breach, compromising the product sterility
Bandage Scissors handle cover component contains an incorrect color masterbatch resin carrier. This may allow a pink dye/residue to appear on the Bandage Scissors and/or sterilization pouches after processing.
Due to software issue, New systems may exhibit an error which may prohibit use of new systems resulting in system being inoperable
An incorrect stopper was used in nine lots of 10mL BD PosiFlush Normal Saline Flush Syringes with General Pump Compatibility. The use of BD PosiFlush Prefilled Flush Syringes with this stopper causes an increase in injection force, which has the potential to trigger alarms if used with a syringe pump.
Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.
Some sensors may have a glucose oxidase (GOX) layer thickness outside of specification (5.5 to 9.2 m). A non-conforming GOX layer thickness could impact sensor function (sensor glucose value) or loss of sensor function due to the sensor layer being too thin. The potential impact resulting from this issue ranges from user inconvenience to over/under delivery of insulin, possibly resulting in hyperglycemia or hypoglycemia, if the sensor glucose values are affected. The issue was identified by the manufacturing team during the execution of Filmetrics Measurement for a Guardian Sensor lot.
BD (C.R. Bard) has received user complaints reporting that when the device was connected to WiFi/Internet, the time displayed on the Sensica device changed during a patient monitoring session which resulted in 1) incorrect time displayed on the monitoring screen, 2) incorrect hour block timestamps on the monitoring screen, 3) incorrect hour timestamps on data screens, and 4) incorrect catheter use time.
New User Manual with additional instructions for use and a hose and mask adapter provided by the device manufacturer. This voluntary recall is intended to help reduce potential health risks of exposure to ozone gas after the use of a SoClean device.
Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
Incorrect product labeling. Exterior product box label does not match internal pouch label.
New User Manual with additional instructions for use and a hose and mask adapter provided by the device manufacturer. This voluntary recall is intended to help reduce potential health risks of exposure to ozone gas after the use of a SoClean device.
VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.
Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.