Medical device and health technology recalls from FDA. Includes diagnostic equipment, implants, monitors, and therapeutic devices.
Potential breach of sterile barrier packaging.
The product Value Assignment Sheet (VAS) is used in calibration of the device. An erroneous value was included in the VAS which has the potential to generate an incorrect APO B patient result which could exceed the total allowable error for the APO B assay (11.6%). The error results in results with a positive bias up to 42%. However, based on the mitigating factors the risk to health is considered to be negligible or limited.
A new software update (v1.20) is introducing error codes for the Dxl 9000 which prevents the assay from being performed and a delay in results from being generated, and may require field service to resolve the delay.
Potential breach of sterile barrier packaging.
Potential breach of sterile barrier packaging.
Firm has received an increase in reports regarding inadvertent instrument lockout during surgical procedures involving the device. If lockout occurs, the device will momentarily activate but will not cut or staple tissue, and additional steps will be required to open it and remove it from tissue. If user is unable to remove the device from the tissue, potential harms to patient include surgical delay, bleeding/hemorrhage, life threatening hemorrhage/hemorrhagic shock, and conversion to open therapy. Two adverse events received, including one patient death.
Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager (DLM) within the LifeShield Medication Management Safety Software and specifies the highest dose at which the pump can be programmed for weight or BSA (body surface area) based medication rulesets. The limit can be configured for Dose, Loading Dose, and Bolus Dose.
Affected kits contain recalled Integra Lifesciences Codman Disposable Perforators due to inadequate welds on specific lots. Perforators with inadequate welds may break or separate.
Software issue with scanning equipment that can results in the filed rotation not being applied correctly.
Notice was issued to clarify flow rates related to the inlet fluid temperature.
Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager (DLM) within the LifeShield Medication Management Safety Software and specifies the highest dose at which the pump can be programmed for weight or BSA (body surface area) based medication rulesets. The limit can be configured for Dose, Loading Dose, and Bolus Dose.
Test results for cast padding component included in procedure trays shows Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals exceeded residual limits, which may lead to a possible loss or lack of functionality that may lead to a delay or prolonged treatment/procedure/therapy.
Due to a manufacturing defect IV catheter may have a hole in the septum, which may result in blood leakage from the septum during insertion, which may cause blood exposure or the need for a second IV to be placed - which may result in therapy delay.
Sampling set nasal cannula used with End-Tidal Carbon Monoxide (ETCO) monitor may have residual carbon monoxide and volatile organic compounds present due to insufficient aeration of the cannula's carbon filter prior to sealing, which may lead to higher than expected ETOCc scores during clinical evaluations.
Due to customer complaints, there is a potential for users to be unable to or find it difficult to remove the plastic plugs in order to inflate the gastric and/or esophageal balloons.
Due to customer complaints, there is a potential for users to be unable to or find it difficult to remove the plastic plugs in order to inflate the gastric and/or esophageal balloons.
Incorrect blister labelling.
Cook Medical identified that devices from the affected device lots may contain the incorrectly sized embolization coil.
Due to customer complaints, there is a potential for users to be unable to or find it difficult to remove the plastic plugs in order to inflate the gastric and/or esophageal balloons.
Cook Medical identified that devices from the affected device lots may contain the incorrectly sized introducer sheath.