DxI 9000 Analyzer (Beckman Coulter) – encoder wheel misalignment (2025)
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137.
Brand
Beckman Coulter, Inc.
Lot Codes / Batch Numbers
UDI-DI: 15099590732103, Serial Numbers: 300116-300117, 300123-300124, 300126, 300128-300139, 300141-300145, 300147- 300158, 300160-300164, 300166-300179, 300181-300183, 300186, 300188-300195, 300197-300207, 300209-300250, 300252- 300256, 300258-300259, 300261, 300263- 300264, 300266-300277, 300279-300290, 300291, 300293-300299, 300300-300309, 300312-300352
Products Sold
UDI-DI: 15099590732103; Serial Numbers: 300116-300117, 300123-300124, 300126, 300128-300139, 300141-300145, 300147- 300158, 300160-300164, 300166-300179, 300181-300183, 300186, 300188-300195, 300197-300207, 300209-300250, 300252- 300256, 300258-300259, 300261, 300263- 300264, 300266-300277, 300279-300290, 300291, 300293-300299, 300300-300309, 300312-300352
Beckman Coulter, Inc. is recalling DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137. due to the DxI 9000 encoder wheel was misaligned on the motor shaft, leading to increased slippage errors and causing the instrument to stop, posing a high s. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
the DxI 9000 encoder wheel was misaligned on the motor shaft, leading to increased slippage errors and causing the instrument to stop, posing a high safety risk due to potential delays in patient results.
Recommended Action
Per FDA guidance
An URGENT MEDICAL DEVICE RECALL notification letter dated 1/30/25 was sent to customers. Affected DxI 9000 Access Immunoassay Analyzers could experience an increase in the following errors: IncubateWheel.Axis MachineControlStatus(Encoder reports that axis position is out of tolerance) WashWheel.Axis MachineControlStatus(Encoder reports that axis position is out of tolerance) This can cause the analyzer to enter the red, or Stopped, system status and any in-process tests are cancelled. There could be a delay in reporting patient test results to the physician. If the error occurs: 1. Confirm that the DxI 9000 Access Immunoassay Analyzer is in either the Paused or Stopped system state. 2. Select the System state button in the lower left corner of the user interface. 3. Select Initialize. When initialization is complete, the analyzer returns to the green, or Running, state. If your laboratory has experienced the issue and it is not resolved by initializing the analyzer, contact your Beckman Coulter representative and schedule a service visit. Beckman Coulter has identified the root cause of this issue and is no longer distributing affected motors for the Incubate and Wash Wheel assemblies. A Beckman Coulter service representative will schedule a visit to replace any affected motors on your DxI 9000 Access Immunoassay Analyzer. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact our Customer Support Center; From our website: http://www.beckmancoulter.com
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.