Becton Dickinson & Company BD 1mL SafetyGlide Allergy and Tuberculin Syringes. Reorder Number 305945 and 305950. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
BD 1mL SafetyGlide Allergy and Tuberculin Syringes. Reorder Number 305945 and 305950.
Brand
Becton Dickinson & Company
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Reorder Number: 305945 Lot: 5312395 and 5312394. Reorder Number: 305950 Lot: 5319508 and 5300175.
Becton Dickinson & Company is recalling BD 1mL SafetyGlide Allergy and Tuberculin Syringes. Reorder Number 305945 and 305950. due to BD received reports of needle assembly disengagement from the syringe.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
BD received reports of needle assembly disengagement from the syringe.
Recommended Action
Per FDA guidance
Recall letters were sent to all wholesalers and end users on 7/28/2006.