Becton Dickinson & Company BD SafetyGlideTM 3ml 25G x 1 3mL SafetyGlide syringes, 50 units in pack Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
BD SafetyGlideTM 3ml 25G x 1 3mL SafetyGlide syringes, 50 units in pack
Brand
Becton Dickinson & Company
Lot Codes / Batch Numbers
Catalog #: 305924 Lot #: 8289599 UDI #: 30382903059240
Products Sold
Catalog #: 305924 Lot #: 8289599 UDI #: 30382903059240
Becton Dickinson & Company is recalling BD SafetyGlideTM 3ml 25G x 1 3mL SafetyGlide syringes, 50 units in pack due to Printed scale markings on the 3mL SafetyGlide syringes, with catalog # 305924, Lot# 8289599, are printed incorrectly. The scale is skewed to different. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Printed scale markings on the 3mL SafetyGlide syringes, with catalog # 305924, Lot# 8289599, are printed incorrectly. The scale is skewed to different degrees, resulting in missing and/or partially printed scale numbers and scale lines.
Recommended Action
Per FDA guidance
BD Sent a customer letter dated 28 July 2020. this letter requests customers to review their inventory for effected lots and Destroy all product subject to the recall. The letter requests customers to return the included response form and notify BD of any adverse events associated to this product.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026