Becton Dickinson & Company BD Vactuainer SST glass and plastic tubes. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
BD Vactuainer SST glass and plastic tubes.
Brand
Becton Dickinson & Company
Lot Codes / Batch Numbers
There is no linkage to any particular lot of BD product.
Products Sold
There is no linkage to any particular lot of BD product.
Becton Dickinson & Company is recalling BD Vactuainer SST glass and plastic tubes. due to "Please be aware tht this not a new recall. The firm has taken action; but, due to administrative issues additional products are now being classified . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
"Please be aware tht this not a new recall. The firm has taken action; but, due to administrative issues additional products are now being classified by the agency". There is an apparent bias in the determination of Total T3, Total T4, HBsAg, Folate BA, VB 12, BR, FSH and Cortisol hormones used to determine thyroid disease states, and adrenal gland function. The immunoassay results are high when
Recommended Action
Per FDA guidance
The firm sent out a Technical Bulletin to all consignees of the specific instrument companies involved. Mailing initial were done 9/17/2004 and a revised mailing was sent 9/23/2004. Global communication will commense 10/5/2004.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026