Becton Dickinson & Company SafeStep Huber Needle Set AllPoints Port Access System; Catalog Numbers AN152075 (UDI 00801741119484), AN152075G (UDI 00801741119583), AN162010 (UDI 00801741119682), AN162010G (UDI 00801741119446), AN162075 (UDI 00801741119675), AN162075G (UDI 00801741119439) Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
SafeStep Huber Needle Set AllPoints Port Access System; Catalog Numbers AN152075 (UDI 00801741119484), AN152075G (UDI 00801741119583), AN162010 (UDI 00801741119682), AN162010G (UDI 00801741119446), AN162075 (UDI 00801741119675), AN162075G (UDI 00801741119439)
Brand
Becton Dickinson & Company
Lot Codes / Batch Numbers
Catalog Numbers AN152075 (UDI 00801741119484), AN152075G (UDI 00801741119583), AN162010 (UDI 00801741119682), AN162010G (UDI 00801741119446), AN162075 (UDI 00801741119675)
Products Sold
Catalog Numbers AN152075 (UDI 00801741119484), AN152075G (UDI 00801741119583), AN162010 (UDI 00801741119682), AN162010G (UDI 00801741119446), AN162075 (UDI 00801741119675), AN162075G (UDI 00801741119439). Lot Numbers: REDT0554 REDW1158 REDW2451 REDX1491 REDY1817 REDY2798 REEN3076 REEP3245 REEP3545 REDR2659 REDS4631 REDT0709 REDT4233 REDU2308 REDV1300 REDV3619 REDX2864 REDY3766 REEP3542 REEQ0644 REER1047 REDX4729 REDZ0700 REDZ1905 REEN1900 REEN2797 REEP3240 REEP4079 REEQ0719 REEQ1251 REER1105 REEQ1250 REDR2870 REDS0199 REDS3479 REDT1592 REDU0734 REDU1349 REDU2234 REDU3360 REDU4418 REDV1066 REDV3684 REDW0388 REDW1159 REDY1595 REDY2936 REDY3545 REDZ1827 REEN2784 REEP0777 REER0610 REER2163 REDS4692 REDT0555 REDU1332 REDU2236 REDV1956 REDW3670 REDX1492 REDX3881 REDX4827 REDY1818 REDY2791 REDY3767 REDZ0388 REDZ1906 REDZ3147 REEN3311 REEQ1172 REER1106 REER2125
Becton Dickinson & Company is recalling SafeStep Huber Needle Set AllPoints Port Access System; Catalog Numbers AN152075 (UDI 00801741119484 due to Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
Recommended Action
Per FDA guidance
URGENT MEDICAL DEVICE RECALL notification letters dated 6/19/20 were sent to customers. PLEASE TAKE THE FOLLOWING ACTIONS: 1. Immediately review your inventory for the specific catalog and lot numbers referenced in Attachment A. Discontinue use of those devices and destroy all product subject to this recall following your institutions process for destruction. MDS-20-1979 Page 2 of 28 BD Medical 1 Becton Drive Franklin Lakes, NJ 07417 USA bd.com 2. Share this notification with all users of the product within your facility network to ensure they are also aware of this recall. 3. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process your credit accordingly for the discarded material. a. Customers who purchased product directly from BD will receive credit. b. Customers who did not purchase product directly from BD will need to contact their distributor to obtain credit. 4. Report any adverse health consequences experienced with the use of this product to BD at product.complaints@bd.com. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1800FDA1088 (18003321088) Mail: MedWatch, HF2, FDA, 5600 Fishers Lane, Rockville, MD 208529787 Customer/Technical Support Phone: 8002901689 Monday Friday between 8:00 am to 5:00 pm MDT SLCBASD.FieldAction@bd.com An Urgent Medical Device Recall notification letter dated 8/25/20 was sent to customers informing them of two additional lots of affected product. An URGENT MEDICAL DEVICE RECALL UPDATE dated 10/7/20 was sent to customers informing them of two additional lots of affected product.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026