Biosense Webster, Inc. CARTO 3 V1.05(FG-5400-00). The CARTO 3 V1.05 System is intended for catheter-based atrial and ventricular mapping. The mapping system allows real-time display of cardiac maps in a number of different formats. Maps may be displayed as cardiac electrical activation maps, cardiac electrical propagation maps, cardiac electrical potential maps, impedance maps and cardiac chamber geometry maps. The acquired patient signals, including body surface ECG and intracardiac electrograms may also be displaye Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
CARTO 3 V1.05(FG-5400-00). The CARTO 3 V1.05 System is intended for catheter-based atrial and ventricular mapping. The mapping system allows real-time display of cardiac maps in a number of different formats. Maps may be displayed as cardiac electrical activation maps, cardiac electrical propagation maps, cardiac electrical potential maps, impedance maps and cardiac chamber geometry maps. The acquired patient signals, including body surface ECG and intracardiac electrograms may also be displaye
Brand
Biosense Webster, Inc.
Lot Codes / Batch Numbers
FG-5400-00
Products Sold
FG-5400-00
Biosense Webster, Inc. is recalling CARTO 3 V1.05(FG-5400-00). The CARTO 3 V1.05 System is intended for catheter-based atrial and ventr due to The recall was initiated because Biosense Webster has become aware of information regarding a potential hazard that may occur during the operation of . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The recall was initiated because Biosense Webster has become aware of information regarding a potential hazard that may occur during the operation of a Bloom EP Stimulator device in conjunction with RF generators and other electrophysiology (EP) equipment, including the CARTO 3 Electropysiology Mapping System.
Recommended Action
Per FDA guidance
Biosense Webster sent an " URGENT FIELD SAFETY NOTICE" dated November 20, 2009 to all affected customers who purchased the Carto 3 Electrophysiology Mapping System. The letter informed the customers of the problem identified and the actions to be taken. Customers were instructed to continue use of the CARTO 3 System and that it was safe and no further action is needed. Customers with questions about the notice were instructed to contact Biosense Webster representatives. On January 12, 2010 another letter was sent from Biosense to all their customers recommending to avoid any connection of an EP stimulator's pacing pins to the direct ports on the CARTO 3 System when applying RF energy, unless the set up is indicated by the stimulator's instructions for use. The letter also informed the customer that the user manual for the CARTO 3 System has been formally revised to include a warning. Customers were instructed to complete an acknowledgement form to confirm receipt of the revised user manual.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.