Covidien Llc Puritan Bennett 980 Series Ventilator, 980A1ENNISB Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Puritan Bennett 980 Series Ventilator, 980A1ENNISB
Brand
Covidien Llc
Lot Codes / Batch Numbers
GTIN/UDI - 10884521201897 Serial Numbers: 35B1700507, 35B1700533, 35B1700539, 35B1700465, 35B1700580, 35B1700530, 35B17005363, 5B1700542, 35B1700509, 35B1701614, 35B1700527, 35B1700537, 35B1700541, 35B1700508 & 35B1701569.
Products Sold
GTIN/UDI - 10884521201897 Serial Numbers: 35B1700507, 35B1700533, 35B1700539, 35B1700465, 35B1700580, 35B1700530, 35B17005363, 5B1700542, 35B1700509, 35B1701614, 35B1700527, 35B1700537, 35B1700541, 35B1700508 & 35B1701569.
Covidien Llc is recalling Puritan Bennett 980 Series Ventilator, 980A1ENNISB due to Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
Recommended Action
Per FDA guidance
On November 4, 2021, the firm began to notify customers through Urgent Medical Device Correction letters. Customers were informed of the device defect that may lead to the ventilator becoming inoperable during use. Actions you should take Immediately discontinue use of the affected ventilators. Remove the affected ventilators from clinical service and quarantine the affected ventilators until such time that a Medtronic Technical Service Engineer inspects and replaces the affected printed circuit board assemblies. Notify all personnel in all care environments in which the affected PB980 series ventilators are used about this medical device correction. If your facility has distributed affected PB980 series ventilators to other persons or facilities, please promptly forward a copy of this letter to those recipients. Complete the attached form and return it as directed to confirm your receipt and understanding of this information. If you are aware of any incidents related to this issue, please contact our Technical Support Department immediately at 1-800-255-6774, option 4, and then option 1 to provide information regarding those events so regulatory reporting obligations can be fulfilled. Work with the Medtronic Technical Support Department if you require assistance finding alternative ventilation devices. Beginning November 5, 2021, Medtronic Technical Service will contact all affected customers to schedule a service appointment for all potentially affected ventilators. If you have any questions regarding this communication, please contact your Medtronic representative or Technical Service at 1-800-255-6774.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026