Diasorin Inc. Liaison 25 OH Vitamin D Assay, Manufacturer part #310900, Lot #114668 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Liaison 25 OH Vitamin D Assay, Manufacturer part #310900, Lot #114668
Brand
Diasorin Inc.
Lot Codes / Batch Numbers
manufacturer part # 310900, Lot #114668
Products Sold
manufacturer part # 310900, Lot #114668
Diasorin Inc. is recalling Liaison 25 OH Vitamin D Assay, Manufacturer part #310900, Lot #114668 due to An internal inspection indicates that Calibrator 2 and Conjugate are switched in their positions in the integral. This will result in Calibrator 1 RL. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
An internal inspection indicates that Calibrator 2 and Conjugate are switched in their positions in the integral. This will result in Calibrator 1 RLU's similar to that of background RLU, and Calibrator 2 RLU's near expected range. A curve can not be calculated from this data.
Recommended Action
Per FDA guidance
DiaSorin Customer Notification dated 02/14/05 was sent to customers instructing them to destroy all kits from the affected lot number. Customers were also asked to complete and return a form indicating how many kits were destroyed.