Diasorin Inc. Version 7 Software for wash program on Bio Tek Automated Microplate Washer Model ELx50 which performs the washing steps for various programmed mictotiter plate assays. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Version 7 Software for wash program on Bio Tek Automated Microplate Washer Model ELx50 which performs the washing steps for various programmed mictotiter plate assays.
Brand
Diasorin Inc.
Lot Codes / Batch Numbers
version 7
Products Sold
version 7
Diasorin Inc. is recalling Version 7 Software for wash program on Bio Tek Automated Microplate Washer Model ELx50 which perform due to When software was updated for the Washer program of the Bio Tek Automated Microplate Washer Model ELx50, the software did not automatically clear old . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
When software was updated for the Washer program of the Bio Tek Automated Microplate Washer Model ELx50, the software did not automatically clear old versions of the wash program when new ones were added.
Recommended Action
Per FDA guidance
Customers were notified by phone and informed of the problem and recommended users to use only the newer wash program which is identified is a 'W' in front of the program.
Verify this information on the official source