Digisonics, Inc Digisonics, Digiview and ERS software releases 3.6.44, 3.6.4.6, and 3.6.5, Digisonics, Inc., 3701 Kirby Drive, Houston, TX 77098, Tel: (800) 940-3240, Fax: (713) 529-7999 Software used to make on-screen (2-D, M-Mode and Doppler) measurements of left ventricle, right ventricle, etc. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Digisonics, Digiview and ERS software releases 3.6.44, 3.6.4.6, and 3.6.5, Digisonics, Inc., 3701 Kirby Drive, Houston, TX 77098, Tel: (800) 940-3240, Fax: (713) 529-7999 Software used to make on-screen (2-D, M-Mode and Doppler) measurements of left ventricle, right ventricle, etc.
Brand
Digisonics, Inc
Lot Codes / Batch Numbers
DigiView software release numbers 3.6.44, 3.6.4.6, and 3.6.5
Products Sold
DigiView software release numbers 3.6.44, 3.6.4.6, and 3.6.5
Digisonics, Inc is recalling Digisonics, Digiview and ERS software releases 3.6.44, 3.6.4.6, and 3.6.5, Digisonics, Inc., 3701 Ki due to Left ventricle volume was inaccurate when the volumes were measured from images having different scales.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Left ventricle volume was inaccurate when the volumes were measured from images having different scales.
Recommended Action
Per FDA guidance
The firm, Digisonics, notified consignees/customers via e-mail entitled "Distribution of Notification of Biplane Error" dated February 6, 2009. The email described the product, problem and actions to be taken. The customers were instructed to immediately contact Digisonics if their site calculates biplane volumes-a utility will be provided which will run on their database and identify all studies in which biplane volumes are calculated. The customers were also instructed to share the following important information with all users of DigiView at their site. The email stated -until the software patch has been installed, biplane volumes should be calculated only from pairs of images having the same depth scale. Replacement DigiView/ERS software that prevents further occurrence of this problem will be provided if it appears their site could be impacted. The Distribution of New Software Release was sent on March 6, 2009. If you have any questions, please call Digisonics Customer Support at 800-344-4266
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026