Software issue
Left ventricle volume was inaccurate when the volumes were measured from images having different scales.
Data from one patient's study had the potential for being put into another patient's report.
Failure of software to update some measurements when tracings or points are amended after measurement calculations have been performed.
Data/measurements from a OB report sent from ultrasound systems populated the wrong OB-View report
Study would on occasion have fetal heart rate and/or cervical length on the report that was not entered nor did it belong to that patient. Software displayed incorrect data for fetal heart rate and cervical length in OB ultrasound reports.