Digisonics, Inc DigiView software release 3.4.4.4 through 3.5.5.2. , 2008 Digisonics, Inc, 800-940-3240. Digiview is a digital image management and reporting system with an associated clinical database. It provides capture, review and storage of images together with report creation and data basing. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
DigiView software release 3.4.4.4 through 3.5.5.2. , 2008 Digisonics, Inc, 800-940-3240. Digiview is a digital image management and reporting system with an associated clinical database. It provides capture, review and storage of images together with report creation and data basing.
Brand
Digisonics, Inc
Lot Codes / Batch Numbers
DigiView software release numbers 3.6.4.4, 3.6.4.6, and 3.6.5.2
Products Sold
DigiView software release numbers 3.6.4.4, 3.6.4.6, and 3.6.5.2
Digisonics, Inc is recalling DigiView software release 3.4.4.4 through 3.5.5.2. , 2008 Digisonics, Inc, 800-940-3240. Digiview i due to Failure of software to update some measurements when tracings or points are amended after measurement calculations have been performed.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Failure of software to update some measurements when tracings or points are amended after measurement calculations have been performed.
Recommended Action
Per FDA guidance
Digisonics, Inc., notified consignees by telephone and e-mail on 10/17/2008, informing them that a problem had been found in the Save Linked Images functionality of DigiView Release 3.6 (and all releases which start with 3.6 such as 3.6.1.). The problem is the failure of the software to update some measurements when tracings or points are amended after measurement calculations have been performed. A utility program, called Linked Image Measurement Verify was being supplied to all sites as a patch to the DigiView program. Release 3.6.6 was to be provided to all customers having previous 3.6 version. Problem was fixed in version 3.6.6 and 3.7.1. If you have any questions, please contact Digisonics Customer Support at (800) 940-3240.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AZ, AR, CA, FL, ID, IL, MD, MA, MI, MO, NY, NC, OH, OK, TX