SIGNA Architect (GE Medical) – Gradient Coil Safety Problem (2025)
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
SIGNA Architect
Brand
GE Medical Systems, LLC
Lot Codes / Batch Numbers
GTIN: 00840682147095 00840682122702 00195278023643 00840682123440
Products Sold
GTIN: 00840682147095 00840682122702 00195278023643 00840682123440
GE Medical Systems, LLC is recalling SIGNA Architect due to GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can p. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.
Recommended Action
Per FDA guidance
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION to its consignees on 05/23/2025 delivered using traceable means. The notice explained the issue, risk, and requested the following: Actions to be taken by Customer/User: You can continue to use your MR system by following the instructions below: Ensure hearing protection with a Noise Reduction Rating (NRR) of at least 33dB is used during exams on these systems. 200 pairs are provided with this letter. Please post this letter in your facility on or near the MR operator console. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form to recall.60999@gehealthcare.com.