Mindray DS USA, Inc. dba Mindray North America ABS800 Chemistry Analyzer Part Number BA81F-PA00001, Containing Gas spring YQ-8/18-90-272-180N (Material Code M6T-010001--- (front spring)) and Gas spring YQ6/15-84-235(B-B)-80N (033-000090-00 (rear spring)) Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
ABS800 Chemistry Analyzer Part Number BA81F-PA00001, Containing Gas spring YQ-8/18-90-272-180N (Material Code M6T-010001--- (front spring)) and Gas spring YQ6/15-84-235(B-B)-80N (033-000090-00 (rear spring))
Brand
Mindray DS USA, Inc. dba Mindray North America
Lot Codes / Batch Numbers
Serial Numbers W2-6C000001, W2-6C000002, W2-6C000003, W2-6C000004, W2-6C000005, XP-5B000611T
Products Sold
Serial Numbers W2-6C000001, W2-6C000002, W2-6C000003, W2-6C000004, W2-6C000005, XP-5B000611T
Mindray DS USA, Inc. dba Mindray North America is recalling ABS800 Chemistry Analyzer Part Number BA81F-PA00001, Containing Gas spring YQ-8/18-90-272-180N (Mate due to The gas spring on the front and rear protective cover may fail. There is a low risk of injury to the user resulting from failure of the gas spring. I. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The gas spring on the front and rear protective cover may fail. There is a low risk of injury to the user resulting from failure of the gas spring. Injury may include impact to the arms, hands or fingers.
Recommended Action
Per FDA guidance
The consignee notification was sent certified mail with return receipt verification. The letter identified the affected product, problem and actions to be taken by the Customer/User. Customers were instructed to: Continue to use the system normally and exercise caution when lifting/closing the front and rear cover. Attachment A includes a list of system(s) identified by model and serial number purchased by your company. We recommend that you notify the clients to which you have distributed these systems and advise them to continue normal use with the recommended caution. The gas springs on the systems listed in Attachment A will require replacement and instructions are provided in Attachment B. The parts needed to perform the replacement will be shipped to you within the next several weeks. Type of Action by the Company Mindray will provide replacement gas springs at no cost. Should you require technical assistance while preforming the replacement, please contact Mindrays Technical Support team at 800-288-2121, option 2 (Technical Support), then choose option 3 (IVD Support team).
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution