The Binding Site Group, Ltd. Optilite Freelite Mx Kappa Free Kit REF LK016.M.OPT.A Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Optilite Freelite Mx Kappa Free Kit REF LK016.M.OPT.A
Brand
The Binding Site Group, Ltd.
Lot Codes / Batch Numbers
Model Number: LK016.M.OPT.A UDI: 05051700019866 Lot Number: 454763
Products Sold
Model Number: LK016.M.OPT.A UDI: 05051700019866 Lot Number: 454763
The Binding Site Group, Ltd. is recalling Optilite Freelite Mx Kappa Free Kit REF LK016.M.OPT.A due to Due to customer complaints of positive bias impacting quality control. The high and low level controls are demonstrating a positive bias compared to . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Due to customer complaints of positive bias impacting quality control. The high and low level controls are demonstrating a positive bias compared to assigned values.
Recommended Action
Per FDA guidance
On 11/30/2020, the firm emailed an "Urgent Field Safety Notice" to its U.S. Customers informing them that investigations have confirmed that the quality control (QC), high and low controls, provided with specific product and lot numbers are consistently demonstrating a ~15% positive bias compared to assigned values and customers may experience QCs consistently reporting results close to the upper acceptable limits, and QC failures may occur on the affected products. Customer are instructed: 1. Users may continue to use any remaining kits of the affected product in their stock. As per the product package insert sample results obtained should only be accepted if the control results are within +/-20% of the concentrations stated on the QC certificate provided. 2. To complete the E-Back Form TSWS18 to Technical.support@bindingsite.com or to their local representative within one week of receiving this notification. Customers are also instructed: -To pass on the notification to all those who need to be aware within their organization or to any organization where the potentially affected devices have been transferred. -To transfer this notification to other organizations on which the information has an impact. -To maintain awareness on this notification for any appropriate period. Any questions, contact the local representative or Technical Support Group: UK technical support is: Telephone: +44(0) 1214569696 Email: technical.support@bindingsite.com Customers in the USA may choose to instead contact the local representative at: Telephone: 800-633-4484 Us-tsc@Bindingsite.com
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
CO, FL, IA, MA, NJ, NY, NC, OH, SD, UT