Loading...
Loading...
All product recalls associated with The Binding Site Group, Ltd..
Total Recalls
15
Past Year
0
Class I (Serious)
0
Most Recent
Oct 2021
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
Due to calibration curves producing a higher activity than routinely expected and that there may be a decrease in separation between calibrator point 5 and calibrator point 6, which can be exacerbated with calibrator 6 activity reaches the optical ceiling of the SPA+ analyzer.
The high and low level "Quality Controls" demonstrate a positive bias of approximately 10% compared to assigned values.
Due to customer complaints of positive bias impacting quality control. The high and low level controls are demonstrating a positive bias compared to assigned values.