The Binding Site Group, Ltd. OPTILITE IgM Kit, REF: NK012.OPT.A, IVD, UDI: 05051700017565 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
OPTILITE IgM Kit, REF: NK012.OPT.A, IVD, UDI: 05051700017565
Brand
The Binding Site Group, Ltd.
Lot Codes / Batch Numbers
Lot # 422251
Products Sold
Lot # 422251
The Binding Site Group, Ltd. is recalling OPTILITE IgM Kit, REF: NK012.OPT.A, IVD, UDI: 05051700017565 due to There is a potential that users may experience difficulty validating the calibration of the kit. This is indicated by kit control values reporting out. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
There is a potential that users may experience difficulty validating the calibration of the kit. This is indicated by kit control values reporting outside the specified ranges. This could result in potential false patient results which could cause a delay until repeat testing is performed.
Recommended Action
Per FDA guidance
On July 29, 2019, the firm issued a "Field Safety Notice" to all consignees. The firm requested consignees do the following: 1. For kit NK012.0PT.A lot 422251, please load reagents on-board the analyser and leave overnight, or alternatively for a minimum of 18 hours, prior to running the calibration and validating the curve. The Optilite analyser holds the assay reagent (R012.0PT lot 446963) in a temperature controlled reagent carousel; please leave the reagents on-board the analyser until empty (following stability advice in the insert Section 6). 2. Customers are advised that once the calibration is established and validated,the reagents must remain on the instrument until the reagent is consumed including when re-running the calibration. Investigations on NK012.0PT.A lot 422251 have shown that inversion of the reagent vials prior to re-calibration can contribute to kit control values reporting outside the specified ranges. 3. Control and calibrator vials are held in the sample handling area of the analyser. When not in use, the control and calibrator vials should be removed from the instrument, be capped to avoid evaporation and be stored at 2-sc in the refrigerator. 4. To validate the calibration curve, the control materials must be tested a minimum of once a day and should return results within 15% of the concentration(s) as stated on the accompanying QC certificate. If a control measurement is out of range when assayed with a stored curve the assay must be recalibrated. If on recalibration the control values measured with the new curve are still out of range, the instrument should be checked before repeating the assay 5. In our laboratories we have found that following the above advice allows continued and safe running of the Optilite lgM assay, if calibration issues are still experienced when following this advice please do not hesitate to contact a Binding Site representative. Advice on action to be taken by the user: " Follow advice
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AL, AZ, AR, CA, FL, GA, IL, IN, IA, KY, MD, MA, MI, MN, NE, NY, NC, OH, OR, PA, SC, SD, TN, TX, UT, VT, WA, WV
Page updated: Jan 10, 2026