The screws may fracture
It was reported that when the GE Precision 500D foot pedal is repeatedly activated, a defect in the fluoro-timer will cause the dose measurement to report a higher value than was actually received by the patient. Patient will receive a lower dosage irradiation time which will result in indication of cumulative air kerma higher than that to which the patient was actually exposed.
Under-Rated Fuses--The Firm may have installed incorrect fuses. The daVinci Surgical System Vision Cart's isolation transformer requires a 10A fuse for 115V operation or a 5A fuse for 230V operation. Operation of this product with the use of an under-rated fuse will result in failure of the fuse and a loss of power to the vision cart, including the loss of power to any ancillary equipment that m