One (1) Fusion Bioline Vascular graft from lot 25162546, which failed guideline straightness testing, was intermingled with lot 25162549 and distributed to customers. Use of the affected graft may result in reduction of blood flow or unwanted tension causing bleeding and the formation of a pseudo-aneurysm.
Direct amplification Discs , used with COVID-19 and Flu A/B & RSV assays, may leak and cause contamination, cross-contamination, invalid results, and error codes.
Siemens Healthineers has received customer complaints and confirmed imprecision for Quality Control (QC) and patient samples with Dimension Tacrolimus (TAC) lots GA2286, GA3047 and GA3171. The health consequences of erroneous patient results are major, since in a worst case scenario, a patient s tacrolimus dosing could be altered, resulting in potential organ failure and rejection (due to too low or no tacrolimus being administered) or toxicity (due to too much tacrolimus being administered). Siemens internal investigation of the Dimension TAC assay showed the worst case imprecision with patient samples at the low end of the Analytical Measurement Range (AMR). A patient sample at 2.0 ng/mL (2.6 nmol/L) recovered as 0.0 ng/mL (0.0 nmol/L) (100% negative bias), another patient sample at 2.1 ng/mL (2.7 nmol/L) recovered at 4.3 ng/mL (5.6 nmol/L) (103% positive bias).