All product recalls associated with Elekta, Inc..
Total Recalls
59
Past Year
2
Class I (Serious)
0
Most Recent
Sep 2025
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
The new "DICOM CT with Body Outline" method to create a patient case in PrecisePLAN 2.15 may produce incorrect results for some customers, given several input conditions encountered in the field. Discontinue use of this method of creating a patient case until FCO 772 00 540 064, (entitled "Update for Patient creation method using DICOM CT with Body Outline"), is received from Elekta to update th
Installation of a touchguard to prevent injury to patients or staff by stopping machine movements in the case of accidental collision.
Possible error in the table longitudinal position.
When using the 'Beam Edit" function to modify a bean in an unlocked Fraction Group, there are two conditions that may cause the user to unitentionally modify a locked fraction group.
PrecisePLAN generates digitally Reconstructed Radiographs (DRR;s) with a shift in the superior direction.
The precision of the calculation used to create the fused study in LGP 8.0 is too low and should not be used until the system is upgraded to LGP 8.2.
After updating the LGK actuator in spare part 810361, the old sleigh became obsolete due to causing insufficient locking of the helmet in combination with the new actuator.
Firm became aware of a failure in the position detection system which could result in incorrect positioning of the detector in the longitudinal or lateral direction.
Under certain conditions it has been found that movement of the collimators can be induced during multi segment beams such as IMRT and Omniwedge which will not be detected by the Desktop Pro if high values are sent within the iCom tolerance table.
Need to modify the closing speed of the shielding doors in the event of an emergency exit
There has been an issue with "Image Fushing" where low precision calculation caused images to become inaccurate.
In some instances, the bolts which are supplied with the table have sheared off, resulting in inability to operate the powered table.
Unexpected Diaphragm movement when manual field size modifications are not saved.
Leksell Gamma Knife C 1.2 with MCU version 4.0.0.6 improperly adjusted trolley for 4mm helmet caused jamming when helmet changer attempted to lock helmet. Lock did not fully engage and could still be raised. When the lock does not engage, then when the helmet changer rises, the helmet could disengage from the helmet changer and fall back onto the helmet trolley.
Due to a material change in a critical component of the linear actuators the quality of said part was found to be inadequate.
Error in dose rates which were calculated at exactly twice the expected values.
Investigation found that if a user accidentally selects the wrong image/tube position images can be displayed flipped.
Customer reported larger than expected variations in wedge factory with gantry rotation.
Images from a different patient can be imported without a warning for the user.