Medical device and health technology recalls from FDA. Includes diagnostic equipment, implants, monitors, and therapeutic devices.
An unexpected shutdown of the IABP may occur due to loss of communication between the Executive Processor PCBA and the Video Generator PCBA. This may result in an interruption to therapy which may threaten the hemodynamic stability of the supported patient as the user is left unaware to the status of the Cardiosave IABP.
There have been reported failures of the high pressure helium regulator, which may cause a helium leak in the Cardiosave Hospital Cart. In instance of helium regulator failure, a Pump Console s internal reservoir of helium will not be replenished when docked into an impacted Hospital Cart. This may result in an insufficient amount of helium within the internal reservoir.
The packaging may be damaged, compromising the sterile barrier.
Select CardioMEMS PA Sensors (Model CM2000) operate outside of the intended frequency range at higher elevations when readings are taken over approximately 2,000 ft. (610 meters) above sea level; operation outside of the intended radiofrequency has the potential to result in inaccurate readings or sensor signal acquisition difficulties.
Radiated Radiofrequency (RF) emissions of the CardioMEMS Patient Electronics Systems (PES) Models CM1000, CM1010 and CM1100 and CardioMEMS Hospital Electronics Systems (HES) Model CM3000 exceed certain emissions limits and have the potential to cause interference with other medical devices, such as implantable neurostimulators, pacemakers, cardiac defibrillators, continuous glucose monitors, or other bio-wearable sensors when in close proximity to an active Hospital or Patient Electronics System (i.e., during readings).
Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted.
Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted.
There have been reported failures of the high pressure helium regulator, which may cause a helium leak in the Cardiosave Hospital Cart. In instance of helium regulator failure, a Pump Console s internal reservoir of helium will not be replenished when docked into an impacted Hospital Cart. This may result in an insufficient amount of helium within the internal reservoir.
An unexpected shutdown of the IABP may occur due to a failure of the connection between the Coiled Cord cable and the Cable Assembly backplane to the Coiled Cord cable which provides the communication between the display head and base unit.
Radiated Radiofrequency (RF) emissions of the CardioMEMS Patient Electronics Systems (PES) Models CM1000, CM1010 and CM1100 and CardioMEMS Hospital Electronics Systems (HES) Model CM3000 exceed certain emissions limits and have the potential to cause interference with other medical devices, such as implantable neurostimulators, pacemakers, cardiac defibrillators, continuous glucose monitors, or other bio-wearable sensors when in close proximity to an active Hospital or Patient Electronics System (i.e., during readings).
An unexpected shutdown of the IABP may occur due to loss of communication between the Executive Processor PCBA and the Video Generator PCBA. This may result in an interruption to therapy which may threaten the hemodynamic stability of the supported patient as the user is left unaware to the status of the Cardiosave IABP.
Stryker has identified a nonconformance in specific lots of Hoffmann LRF Bone Transport Struts. Specifically, the thread pitch is oversized at the distal end of the threaded rod, and as a result the distal ring fixation cannot be locked between 0- 50mm of the threaded rod.
Sedecal was informed of 3 incidents of erratic movements related to X-ray mobile systems.
The mobile application for Android devices downloaded from the Google Play Store do not load per specifications and function on the Android devices.
The outer package labeling and product etch are a 40 mm Size D liner, however, the implant is a 38 mm Size C liner.
For users with MYLA V4.8.X / V4.9 that use VITEK MS to identify organisms as part of their workflow, AST filter rules that have been activated are not always being applied to AST results when sending the results to the clinician.
Sedecal was informed of 3 incidents of erratic movements related to X-ray mobile systems.
Risk of boom pivot failing due to the nut securing the boom becoming loose and the bolt to be unsecure to the flange on the side.
Removes the test definition for HemosIL Liquid Anti-Xa (PN 0020302600 and 0020302601) from the ACL Elite/Elite Pro instruments
Risk of boom pivot failing due to the nut securing the boom becoming loose and the bolt to be unsecure to the flange on the side.