Medical device and health technology recalls from FDA. Includes diagnostic equipment, implants, monitors, and therapeutic devices.
Risk of boom pivot failing due to the nut securing the boom becoming loose and the bolt to be unsecure to the flange on the side.
Removes the test definition for HemosIL Liquid Anti-Xa (PN 0020302600 and 0020302601) from the ACL Elite/Elite Pro instruments
Removes the test definition for HemosIL Liquid Anti-Xa (PN 0020302600 and 0020302601) from the ACL Elite/Elite Pro instruments
Under specific circumstances, the U.S. product registration system did not send up enablement requests to U.S. LATITUDE Customer Support during an 8-month interval in 2022. For these patients, the HeartLogic Index and any associated Yellow Alerts are currently not visible in LATITUDE as may have been intended by the healthcare provider..
The field action was initiated in February of 2023 and was limited to the affected lot from sales force trunk stock only.
In the event of any unintended table movement, the system may not detect the incorrect direction, could lead to the injury of a patient, staff member, operator, or equipment.
Stryker has identified an issue that impacts two specific lots of CLAW" II plates. The plates within these two lots were found to have been comingled, resulting in the incorrect product in packages. Packaging for 402S0425 incorrectly contained 40241425, and vice versa.
Cease use of the Habib EndoHPB Bipolar RF Catheter with the ERBE VIO3 Generator immediately due to the potential for delivery of excessive energy leading to thermal injury or tissue damage. An inaccurate setting is currently provided in the HABIB EndoHPB Catheter's Instructions for Use ("IFU").
Product is labeled with an incorrect expiration date.
Microcatheters for embolic coil and diagnostic agent introduction to peripheral/neuro vasculature, due to manufacturing non-conformance, and due to flush, guidewire insertion and manipulation prior to placement, may have loose particles within catheter lumens that may be introduced to patients, which could cause embolic event, small vessel occlusion, and subacute to chronic inflammatory response.
MiniCap Disconnect Cap with Povidone-Iodine are packaged in foil pouches, which may have been incorrectly sealed, i.e., the pouches may have open or weak seals. This could lead to exposure to air, resulting in insufficient iodine/dry sponge inside the MiniCap, which could lead to the potential for inadequate disinfectant.
Transport Cot has insufficient volume of epoxy to assemblies installed in the head section and foot sectioning system may result in separation of head end or foot end from the cot and result in unexpected disengagement while loading/unloading or transport.
New warnings and precautionary statements on product labelling: Reagent 2 (R2) of BIOPHEN Protein C 5 contains the chemical substance - Cesium Chloride which changed classification and has become hazardous according to the CLP regulation ((EC) No 1272/2008) is suspected of damaging fertility
Nonsterile syringe potentially exposing the patient to pathogenic microorganisms and a risk of local infection.
Due to temperature excursions with products (products being stored outside of the required temperature specifications) that may cause a reduction in performance. The temperature excursion was as a result of shipping delays.
Automated external defibrillators were potentially distributed with the incorrect language configuration, which can lead to a delay in therapy or no therapy.
TFNA Femoral Nails labeled and etched as 200mm may measure 170mm in length, and TFNA Femoral Nails labeled and etched as 170mm may measure 200mm in length
Nonsterile syringe potentially exposing the patient to pathogenic microorganisms and a risk of local infection.
Kits with affected microtiter plates have the potential to produce low OD values, causing the run to fail, per the Validation requirements of the Package Insert.
The Antibody (R2 reagent) is showing a positive bias compared to previous batches on patient samples. Quality Controls do not show the same bias and will be within range