Draeger Medical Systems, Inc. Draeger Infinity M300 and M300+, for use with the ICS to monitor ECG and pulse oximetry on ambulatory and non-ambulatory adult and pediatric patients using wireless communication over the Infinity patient monitoring network, Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Draeger Infinity M300 and M300+, for use with the ICS to monitor ECG and pulse oximetry on ambulatory and non-ambulatory adult and pediatric patients using wireless communication over the Infinity patient monitoring network,
Brand
Draeger Medical Systems, Inc.
Lot Codes / Batch Numbers
UDI-DI: 04049098048989 (M300), 04049098095778 (M300+), All Systems Running Software versions: VG2.4 and lower
Products Sold
UDI-DI: 04049098048989 (M300), 04049098095778 (M300+); All Systems Running Software versions: VG2.4 and lower
Draeger Medical Systems, Inc. is recalling Draeger Infinity M300 and M300+, for use with the ICS to monitor ECG and pulse oximetry on ambulator due to The software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response r. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this condition, ECG waveforms with unusually narrow and/or high frequency QRS may be displayed or printed with QRS amplitudes intermittently lower than actual.
Recommended Action
Per FDA guidance
An Urgent - Medical Device Correction notification letter dated March 2023 was sent to customers. Actions to be taken: Until the corrected software is available and installed, clinicians should utilize the Draeger ICS/MSOO Instructions For Use while applying any additional means of standard practices for assessment of patients with unusually high and/or low QRS amplitudes. Clinicians should combine this information with physical assessment of the patient and follow standard clinical practice. Please ensure that all users of the affected products listed above and other people within your organization are made aware of this Important Safety Notice. If you have provided the products to third parties, please forward a copy of this information to them. Actions being taken by the Company: We are currently working on releasing updated software (ICS VG4/M300 VG3) to resolve the issue which will be provided to you at no charge; targeted for the third quarter of 2023. Your local Drager Service representative will contact you to arrange a date for the software update to be performed once the software is available. Please keep this information at least until the software update has been completed. Please also complete and return the Customer Acknowledgment Card provided to confirm to us that you have received this information. We regret any inconvenience this may cause. We consider this notice a necessary measure to increase patient safety. We thank you for your support. If you have any questions regarding the operation of the Product, please contact Drager Service Technical Support between the hours of 8:00 AM - 8:00 PM EST at 1-800-437-2437 (press 2 at the prompt, then 2, then 1). An Urgent - Medical Device Correction UPDATE TO: Urgent - Medical Device Correction released March 2023 relating to software on the Infinity CentralStation dropping peaks on narrow waveforms causing the Infinity M300 to fail clauses 201.12.1.101 and 201.12.1.101.8 within IEC 60601-2-27 Frequ
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026