Draeger Medical Systems, Inc. Infinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, Model No. MS20407. Rev 20, 21, and 23 Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Infinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, Model No. MS20407. Rev 20, 21, and 23
Brand
Draeger Medical Systems, Inc.
Lot Codes / Batch Numbers
Model Number: MS20407. UDI-DI: 04049098054447. Rev 20, 21, and 23
Products Sold
Model Number: MS20407. UDI-DI: 04049098054447. Rev 20, 21, and 23
Draeger Medical Systems, Inc. is recalling Infinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, Model No. MS20407. Re due to Device is not in full compliance to Type CF requirements of IEC 60601-1 and IEC 60601-2-34 standards. Therefore, there is a potential risk of electric. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Device is not in full compliance to Type CF requirements of IEC 60601-1 and IEC 60601-2-34 standards. Therefore, there is a potential risk of electric shock. Health consequences may include electrical shock, burns, and cardiac arrhythmia.
Recommended Action
Per FDA guidance
On May 13, 2024, the firm began notifying affected customers of the recall through letters titled "Urgent - Medical Device Correction". Customers were informed of the issue and given additional precautions to inspect the patient cables and probes, and to discard damaged or compromised accessories. Customers will be contacted by their local Draeger Service representative to arrange for modification of the affected M500 units to be carried out free of charge as soon as the material for modification is available. If you have any questions regarding the operation of the Product, please contact Drager Service Technical Support between the hours of 8:00 AM - 8:00 PM EST at 1-800-437-2437 (press 2 at the prompt, then 2, then 1). If you have any questions regarding this letter, please contact Michael Kelhart between the hours of 8:00 AM -4:30 PM EST at 267-664-1131 or via email at mike.kelhart@draeger.com.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.