Draeger Medical Systems, Inc. Draeger Softbed Resuscitaire and/or Birthing Room Warmer with Scale [Model(s): RW82VHA-1, RW82VHA-1C, RW82VHA-1D, RW82-1, RW82-1C, RW82-1D, WBR82-1 and WBR82-1C] Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Draeger Softbed Resuscitaire and/or Birthing Room Warmer with Scale [Model(s): RW82VHA-1, RW82VHA-1C, RW82VHA-1D, RW82-1, RW82-1C, RW82-1D, WBR82-1 and WBR82-1C]
Brand
Draeger Medical Systems, Inc.
Lot Codes / Batch Numbers
All units shipped from January 20, 2020 through March 1, 2023
Products Sold
All units shipped from January 20, 2020 through March 1, 2023; MU20503 (Resuscitaire) UDI - 04049098000215 MU20512 (Birthing Room Warmer - BRW) UDI - 04049098058001
Draeger Medical Systems, Inc. is recalling Draeger Softbed Resuscitaire and/or Birthing Room Warmer with Scale [Model(s): RW82VHA-1, RW82VHA-1C due to The Resuscitaire Infant Radiant Warmer with the optional scale could display inaccurate weight values.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The Resuscitaire Infant Radiant Warmer with the optional scale could display inaccurate weight values.
Recommended Action
Per FDA guidance
An Urgent - Medical Device Correction dated May 2023 was sent to customers. Actions to be taken: Please ensure that all users of the above-mentioned products and other persons within your organization are made aware of this Important Safety Notice. If you have provided the products to third parties, please forward a copy of this information. Users can mitigate this risk by performing the visual check, test procedure and take appropriate actions based on the Attachment provided in this notice. Complete and return the enclosed Customer Acknowledgement and Response Form . We are continuing to evaluate this issue and will advise the affected customers when we have an alternative solution. Visual Check for Correct Mattress Placement: To ensure proper operation of the scale, please check to confirm there is a clearance between the mattress cover and the corner blocks and side panels of the Resuscitaire. 1. Ensure mattress with cover is installed in the Resuscitaire bassinet. 2. Observe edges of mattress for clearance with corner blocks and side panels, see Figure 2. 3. If there is interference, then the scale may not provide accurate weight measurements. It is recommended to use an independent scale for therapeutic decisions and treatment Test Procedure If it is unclear from the visual check if the mattress cover interferes with the corner blocks and/or side panels, impacting scale function, please perform the following test: a. Place a known calibrated weight on the center mattress (with mattress cover installed). A 5kg weight is recommended in the instructions for use. b. Execute weighing procedure and record weight. c. Repeat weighing procedure 5-10 times. d. Compare measurements. i. If measurements are within 5 grams of calibrated 5kg weight, mattress assembly is acceptable. ii. If measurements deviate more than +/- 5 grams from the calibrated 5kg weight, it is recommended to use an alternate weighing scale Contact: We regret any inconvenience this may cause.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.