Hamilton Co 300 Microliter CO-RE Tips for use with Microlab STAR, manufactured by Hamilton Company, Reno, NV. Product is a disposable product used with Microlab STAR series Instruments, which are microtiter diluting and dispensing devices. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
300 Microliter CO-RE Tips for use with Microlab STAR, manufactured by Hamilton Company, Reno, NV. Product is a disposable product used with Microlab STAR series Instruments, which are microtiter diluting and dispensing devices.
Brand
Hamilton Co
Lot Codes / Batch Numbers
Part Number 235950 (case containing 11, 520 tips), Lot Number 2950441 and Part Number 235902 (case containing 5, 760 tips), Lot Number 2950461.
Products Sold
Part Number 235950 (case containing 11,520 tips), Lot Number 2950441 and Part Number 235902 (case containing 5,760 tips), Lot Number 2950461.
Hamilton Co is recalling 300 Microliter CO-RE Tips for use with Microlab STAR, manufactured by Hamilton Company, Reno, NV. P due to Product Defect-- some tips from the mold cavity 28 are too short and do not always aspirate or dispense liquid correctly.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Product Defect-- some tips from the mold cavity 28 are too short and do not always aspirate or dispense liquid correctly.
Recommended Action
Per FDA guidance
The firm initiated the recall on April 10, 2009, and issued notification to consignees via certified mail and e-mail. Customers were instructed to return the affected product to the Hamilton Company. A fax-back form was supplied with each notification. Direct questions about the recall to the Hamilton Company by calling 1-775-858-3000, extension 236.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
CA, DE, MA, NJ, PA