All product recalls associated with Hamilton Co.
Total Recalls
11
Past Year
0
Class I (Serious)
0
Most Recent
Jan 2021
Data from FDA, USDA, CPSC, and NHTSA public records. Recall counts vary by company size, industry, and reporting practices. Always check official sources for current information.
There is a risk that the device can start a fire if flammable liquids contact and enter the device.
Product Defect-- some tips from the mold cavity 28 are too short and do not always aspirate or dispense liquid correctly.
Some tips may be occluded and are unable to aspirate or dispense liquid.
Manufacturing issue with spring may restrict product to drawing less than 1/2 of its stated measured capacity
Incorrectly labeled-- Product has the wrong barcode labeling. The barcode error will not influence the results of testing. The instrument will discover the error and post an error notice.
Some product containers may be leaking.
Some of the packaged tips have increased force during aspiration, which can lead to pipetting errors such as LLD error messages, and also has risk of causing a missed pipetting.
There is an error in the marking of the graduated divisions on syringes in affected lots that shows an amount that is double the actual volume of the syringe
There is an error in the marking of the graduated divisions on syringes in affected lots that shows an amount that is double the actual volume of the syringe
Heat sinking on power boards may not operate as intended and the board may cease to operate.
The product valve has a high probability of deforming and may result in leakage during use.