Hamilton Co Hamilton brand Disposable Precision Tips, Part Number 235300 (box containing 504 tips) and 235400 (case containing 20 boxes), Manufactured by Hamilton Company, Reno, Nevada Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Hamilton brand Disposable Precision Tips, Part Number 235300 (box containing 504 tips) and 235400 (case containing 20 boxes), Manufactured by Hamilton Company, Reno, Nevada
Brand
Hamilton Co
Lot Codes / Batch Numbers
Lot: 33906 P5 and Lot: 63906 P5
Products Sold
Lot: 33906 P5 and Lot: 63906 P5
Hamilton Co is recalling Hamilton brand Disposable Precision Tips, Part Number 235300 (box containing 504 tips) and 235400 (c due to Some of the packaged tips have increased force during aspiration, which can lead to pipetting errors such as LLD error messages, and also has risk of . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Some of the packaged tips have increased force during aspiration, which can lead to pipetting errors such as LLD error messages, and also has risk of causing a missed pipetting.
Recommended Action
Per FDA guidance
The firm initiated its recall on January 30, 2007. The initial notification was sent via certified mail and e-mail. Subrecalls will be conducted by the distributor.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026