Hamilton Co Microlab AT, part numbers 135001, 135001OOC, 135002, 135003, 135005, 135006, 135006R, 135007, 135008, 135016, 135028, 135029, 936400, 936400BLU. Microlab AT+ and Microlab AT+2, part numbers 142000, 142001, 142001R, 142002, 142003, 142005, 142006, 142006R, 142007, 142008, 142009, 142015, 142017, 142035, 142036 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Microlab AT, part numbers 135001, 135001OOC, 135002, 135003, 135005, 135006, 135006R, 135007, 135008, 135016, 135028, 135029, 936400, 936400BLU. Microlab AT+ and Microlab AT+2, part numbers 142000, 142001, 142001R, 142002, 142003, 142005, 142006, 142006R, 142007, 142008, 142009, 142015, 142017, 142035, 142036
Brand
Hamilton Co
Lot Codes / Batch Numbers
All codes included.
Products Sold
All codes included.
Hamilton Co is recalling Microlab AT, part numbers 135001, 135001OOC, 135002, 135003, 135005, 135006, 135006R, 135007, 135008 due to Heat sinking on power boards may not operate as intended and the board may cease to operate.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Heat sinking on power boards may not operate as intended and the board may cease to operate.
Recommended Action
Per FDA guidance
On 12/15/05, the firm initiated the recall and its notification was via letters explaining the reason for the recall and requesting the product be returned.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
IN