Hamilton Co Syringe, 62RNR 2.5 Microliter Syringe, with needle point style 3 (blunt needle point for use with HPLC injection valves and for sample pipetting) Part #: 87942. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Syringe, 62RNR 2.5 Microliter Syringe, with needle point style 3 (blunt needle point for use with HPLC injection valves and for sample pipetting) Part #: 87942.
Brand
Hamilton Co
Lot Codes / Batch Numbers
Lot 340007 and 342081
Products Sold
Lot 340007 and 342081
Hamilton Co is recalling Syringe, 62RNR 2.5 Microliter Syringe, with needle point style 3 (blunt needle point for use with HP due to There is an error in the marking of the graduated divisions on syringes in affected lots that shows an amount that is double the actual volume of the . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
There is an error in the marking of the graduated divisions on syringes in affected lots that shows an amount that is double the actual volume of the syringe
Recommended Action
Per FDA guidance
On 3/28/06, the firm initiated the recall and its notification was via letters explaining the reason for the recall and requesting the product be returned.