Draeger Medical Systems, Inc. Stabilet Infant Warming System, model numbers 200, 300, 1250, 1500, 200/3000, 2000, 2200/3200, 3000, and 3200. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Stabilet Infant Warming System, model numbers 200, 300, 1250, 1500, 200/3000, 2000, 2200/3200, 3000, and 3200.
Brand
Draeger Medical Systems, Inc.
Lot Codes / Batch Numbers
All units of model numbers 200, 300, 1250, 1500, 200/3000, 2000, 2200/3200, 3000, and 3200.
Products Sold
All units of model numbers 200, 300, 1250, 1500, 200/3000, 2000, 2200/3200, 3000, and 3200.
Draeger Medical Systems, Inc. is recalling Stabilet Infant Warming System, model numbers 200, 300, 1250, 1500, 200/3000, 2000, 2200/3200, 3000, due to Hot particles fell into the infant bassinet beneath a Radiant Infant Warmer. The resulting fire injured an infant. These devices are obsolete.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Hot particles fell into the infant bassinet beneath a Radiant Infant Warmer. The resulting fire injured an infant. These devices are obsolete.
Recommended Action
Per FDA guidance
On 7/20/09, Draeger Medical, the recalling firm, issued a press release to medical media outreach organizations and an Urgent Medical Device letter to all known customers. This information explained the incident and the need to remove and replace the units in question as soon as feasibly possible.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026